Back to Search Start Over

Variability of transient elastography-based spleen stiffness performed at 100 Hz.

Authors :
Armandi A
Merizian T
Werner MM
Coxson HO
Sanavia T
Birolo G
Gashaw I
Ertle J
Michel M
Galle PR
Labenz C
Emrich T
Schattenberg JM
Source :
European radiology experimental [Eur Radiol Exp] 2023 Dec 12; Vol. 7 (1), pp. 79. Date of Electronic Publication: 2023 Dec 12.
Publication Year :
2023

Abstract

Background: Spleen stiffness measurement (SSM) performed by transient elastography at 100 Hz is a novel technology for the evaluation of portal hypertension in advanced chronic liver disease, but technical aspects are lacking. We aimed to evaluate the intraexamination variability of SSM and to determine the best transient elastography protocol for obtaining robust measurements to be used in clinical practice.<br />Methods: We analyzed 253 SSM exams with up to 20 scans for each examination, performed between April 2021 and June 2022. All SSM results were evaluated according to different protocols by dividing data into groups of n measurements (from 2 to 19). Considering as reference the median SSM values across all the 20 measurements, we calculated the distribution of the absolute deviations of each protocol from the reference median. This analysis was repeated 1,000 times by resampling the data. Distributions were also stratified by etiology (chronic liver disease versus clinically significant portal hypertension) and different SSM ranges: < 25 kPa, 25-75, and > 75 kPa.<br />Results: Overall, we observed that the spleen stiffness exam had less variability if it exceeded 12 measurements, i.e., absolute deviations ≤ 5 kPa at 95% confidence. For exams with higher SSM values (> 75 kPa), as seen in clinically significant portal hypertension, at least 15 measurements are highly recommendable.<br />Conclusions: Fifteen scans per examination should be considered for each SSM exam performed at 100 Hz to achieve a low intraexamination variability within a reasonable time in clinical practice.<br />Relevance Statement: Performing at least 15 scans per examination is recommended for 100 Hz SSM in order to achieve a low intraexamination variability, in particular for values > 75 kPa compatible with clinically significant portal hypertension.<br />Key Points: • Spleen stiffness measurement by transient elastography is used for stratification in patients with portal hypertension. • At 100 Hz, this method may have intraexamination variability. • A minimum of 15 scans per examination achieves a low intraexamination variability.<br /> (© 2023. The Author(s).)

Details

Language :
English
ISSN :
2509-9280
Volume :
7
Issue :
1
Database :
MEDLINE
Journal :
European radiology experimental
Publication Type :
Academic Journal
Accession number :
38087079
Full Text :
https://doi.org/10.1186/s41747-023-00393-2