Comparisons of biometric parameters measurements by OPD-SCAN Ⅲ and Pentacam in cataract.
- Source :
-
Photodiagnosis and photodynamic therapy [Photodiagnosis Photodyn Ther] 2024 Feb; Vol. 45, pp. 103911. Date of Electronic Publication: 2023 Dec 02. - Publication Year :
- 2024
-
Abstract
- Background: Comparison of OPD-Scan III and Pentacam measurements of relevant parameters guiding the selection of multifocal intraocular lenses in cataract patients.<br />Methods: A retrospective study included 120 patients (120 eyes) with cataract. Their chord kappa, chord alpha, corneal spherical aberration, and pupil size were measured by OPD-Scan Ⅲ and Pentacam. Pairwise t-tests, Pearson tests, and Bland-Altman analyses were used, respectively, to assess the difference, correlation, and agreement between the devices.<br />Results: There was no significant difference in photopic chord kappa and corneal spherical aberration between the two instruments (P = 0.054, P = 0.065). Chord alpha and pupil size varied significantly between the two instruments (P < 0.001). OPD-Scan III results revealed a significant association between photopic chord kappa and mesopic chord kappa (r = 0.823, P < 0.001). There was a significant positive correlation between photopic and mesopic chord kappa measured by OPD-Scan Ⅲ and chord kappa measured by Pentacam (r = 0.840, r  = 0.757, P < 0.001). The chord alpha evaluated by the two tools had a moderate correlation (r = 0.442, P < 0.001) between them. Bland-Altman analysis showed that there was excellent agreement between the two instruments in measuring chord kappa and corneal spherical aberration, while the chord alpha and pupil size were not consistent.<br />Conclusions: The chord kappa and corneal spherical aberration measured by OPD-Scan III and Pentacam were consistent in the preoperative decision making of multifocal intraocular lens implantation in cataract patients, both of which have clinical guiding significance. In addition, OPD-Scan III provides more comprehensive day and night chord kappa and pupil diameters, which improves clinical advice for patients with complicated ocular diseases and high demands for nocturnal vision.<br />Competing Interests: Declaration of Competing Interest The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.<br /> (Copyright © 2024 The Authors. Published by Elsevier B.V. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1873-1597
- Volume :
- 45
- Database :
- MEDLINE
- Journal :
- Photodiagnosis and photodynamic therapy
- Publication Type :
- Academic Journal
- Accession number :
- 38043761
- Full Text :
- https://doi.org/10.1016/j.pdpdt.2023.103911