A singlicate immunogenicity method to detect anti-polyethylene glycol antibodies: pre- and post-dose of PEGylated therapies.

Authors :
Jyamubandi I
Aamir A
Kaur J
Sylvester A
Hughes R
Li S
Source :
Bioanalysis [Bioanalysis] 2024 Jan; Vol. 16 (1), pp. 5-17. Date of Electronic Publication: 2023 Nov 22.
Publication Year :
2024

Abstract

Aim: Preexisting anti-polyethylene glycol (PEG) antibodies (APAs) may affect the efficacy and safety of PEGylated compounds. Omontys <superscript>®</superscript> and Krystexxa <superscript>®</superscript> withdrawal and SARS-CoV-2 RNA vaccine anaphylaxis have all been linked to APAs. This project aimed to develop and validate a method to detect total antibodies against PEG, pre- and post-dose. Materials & methods: The repetitive, linear PEG structure prevented the use of a bridging homogenous format, hence the requirement to use a solid-phase extraction and acid dissociation assay coupled with the Meso Scale Discovery <superscript>®</superscript> platform. Results & conclusion: Using singlicate analysis, the method was validated to successfully detect APA pre- and post-dose, with a crucial aspect of the method being the preparation of an appropriate negative control.