Back to Search Start Over

Outcomes and performance of the ACURATE neo2 transcatheter heart valve in clinical practice: one-yearresults of the ACURATE neo2 PMCF Study.

Authors :
Kim WK
Möllmann H
Montorfano M
Ellert-Gregersen J
Rudolph TK
Van Mieghem NM
Hilker M
Amat-Santos I
Terkelsen CJ
Petronio AS
Stella P
Götberg M
Rück A
Kasel AM
Trillo R
Appleby C
Barbanti M
Blanke P
Asch FM
Modolo R
Allocco DJ
Tamburino C
Source :
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology [EuroIntervention] 2024 Jan 01; Vol. 20 (1), pp. 85-94.
Publication Year :
2024

Abstract

Background: Transcatheter aortic valve implantation is an effective treatment for patients with aortic stenosis; however, complications related to paravalvular leakage (PVL) persist, including increased risk of mortality, cardiovascular mortality, and rehospitalisation.<br />Aims: We sought to evaluate the clinical outcomes and valve performance at 1 year in patients with severe aortic stenosis treated with the ACURATE neo2 valve in a post-market clinical setting.<br />Methods: Valve Academic Research Consortium-2 safety events were assessed up to 1 year. Independent core laboratories evaluated echocardiographic measures of valve performance and hypoattenuated leaflet thickening (HALT; as measured by four-dimensional computed tomography).<br />Results: The study enrolled 250 patients (64% female; mean age: 81 years; baseline Society of Thoracic Surgeons risk score: 2.9±2.0%); 246 patients were implanted with ACURATE neo2. All-cause mortality was 0.8% at 30 days and 5.1% at 1 year. The 1-year rates for stroke and disabling stroke were 3.0% and 1.3%, respectively. Overall, HALT of >50% leaflet involvement of at least one leaflet was present in 9% of patients at 30 days and in 12% of patients at 1 year. No association was observed between the presence of HALT and 1-year clinical or haemodynamic outcomes. Early haemodynamic improvements were maintained up to 1 year (mean aortic valve gradient: 47.6±14.5 mmHg at baseline, 7.6±3.2 mmHg at 1 year; mean aortic valve area: 0.7±0.2 cm <superscript>2</superscript> at baseline, 1.7±0.4 cm <superscript>2</superscript> at 1 year). At 1 year, 99% of patients had mild or no/trace PVL (<1% had moderate PVL; no patient had severe PVL).<br />Conclusions: The study outcomes confirm favourable performance and safety up to 1 year in patients treated with ACURATE neo2 in routine clinical practice. (ClinicalTrials.gov: NCT04655248).

Details

Language :
English
ISSN :
1969-6213
Volume :
20
Issue :
1
Database :
MEDLINE
Journal :
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
37982152
Full Text :
https://doi.org/10.4244/EIJ-D-23-00823