Back to Search Start Over

Editor's Choice - Radiation Dose Reduction During Contralateral Limb Cannulation Using Fiber Optic RealShape Technology in Endovascular Aneurysm Repair.

Authors :
Klaassen J
Hazenberg CEVB
Bloemert-Tuin T
Wulms SCA
Teraa M
van Herwaarden JA
Source :
European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery [Eur J Vasc Endovasc Surg] 2024 Apr; Vol. 67 (4), pp. 594-600. Date of Electronic Publication: 2023 Nov 03.
Publication Year :
2024

Abstract

Objective: The increasing number of endovascular procedures has resulted in an increasing radiation burden, particularly for the treatment team. Fiber Optic RealShape (FORS) technology uses laser light instead of fluoroscopy to visualise the endovascular guidewire and catheters. These devices can be used during the navigational part of procedures, such as cannulation of the contralateral limb (CL) in endovascular aneurysm repair (EVAR). The aim of this study was to describe the effect of using FORS on radiation dose during CL cannulation in standard EVAR.<br />Methods: This was a non-randomised, retrospective comparison study of prospectively collected, single centre data from FORS guided EVAR compared with a conventional fluoroscopy only guided EVAR cohort. A total of 27 FORS guided cases were matched 1:1 based on sex, age, and body mass index (BMI) with 27 regular (fluoroscopy only) EVARs. This study primarily focused on (1) technical success of FORS and (2) navigation time and radiation dose (cumulative air kerma [CAK], air kerma area product [KAP], and fluoroscopy time [FT]) during cannulation of the CL. In addition, overall procedure time and radiation dose of the complete EVAR procedure were studied.<br />Results: In 22 (81%) of the 27 FORS guided cases the CL was successfully cannulated using FORS. All radiation dose parameters were significantly lower in the FORS group (CAK, p < .001; KAP, p = .009; and FT, p < .001) for an equal navigation time (p = .95). No significant differences were found when comparing outcomes of the complete procedure.<br />Conclusion: Use of FORS technology significantly reduces radiation doses during cannulation of the CL in standard EVAR.<br /> (Copyright © 2023 The Authors. Published by Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1532-2165
Volume :
67
Issue :
4
Database :
MEDLINE
Journal :
European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery
Publication Type :
Academic Journal
Accession number :
37925100
Full Text :
https://doi.org/10.1016/j.ejvs.2023.10.041