Back to Search
Start Over
Safety and performance of a hybrid closed-loop insulin delivery system with carbohydrate suggestion in adults with type 1 diabetes prone to hypoglycemia.
- Source :
-
Diabetes research and clinical practice [Diabetes Res Clin Pract] 2023 Nov; Vol. 205, pp. 110956. Date of Electronic Publication: 2023 Oct 14. - Publication Year :
- 2023
-
Abstract
- Aims: To evaluate the safety and performance of a hybrid closed-loop (HCL) system with automatic carbohydrate suggestion in adults with type 1 diabetes (T1D) prone to hypoglycemia.<br />Methods: A 32-hour in-hospital pilot study, including a night period, 4 meals and 2 vigorous unannounced 45-minute aerobic sessions, was conducted in 11 adults with T1D prone to hypoglycemia. The primary outcome was the percentage of time in range 70-180 mg/dL (TIR). Main secondary outcomes were time below range < 70 mg/dL (TBR < 70) and < 54 (TBR < 54). Data are presented as median (10th-90th percentile ranges).<br />Results: The participants, 6 (54.5%) men, were 24 (22-48) years old, and had 22 (9-32) years of T1D duration. All of them regularly used an insulin pump and a continuous glucose monitoring system. The median TIR was 78.7% (75.6-91.2): 92.7% (68.2-100.0) during exercise and recovery period, 79.3% (34.9-100.0) during postprandial period, and 95.4% (66.4-100.0) during overnight period. The TBR < 70 and TBR < 54 were 0.0% (0.0-6.6) and 0.0% (0.0-1.2), respectively. A total of 4 (3-9) 15-g carbohydrate suggestions were administered per person. No severe acute complications occurred during the study.<br />Conclusions: The HCL system with automatic carbohydrate suggestion performed well and was safe in this population during challenging conditions in a hospital setting.<br />Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (Copyright © 2023 Elsevier B.V. All rights reserved.)
- Subjects :
- Male
Adult
Humans
Young Adult
Middle Aged
Female
Insulin adverse effects
Blood Glucose
Hypoglycemic Agents adverse effects
Blood Glucose Self-Monitoring
Pilot Projects
Treatment Outcome
Insulin Infusion Systems
Insulin, Regular, Human therapeutic use
Diabetes Mellitus, Type 1 drug therapy
Hypoglycemia epidemiology
Subjects
Details
- Language :
- English
- ISSN :
- 1872-8227
- Volume :
- 205
- Database :
- MEDLINE
- Journal :
- Diabetes research and clinical practice
- Publication Type :
- Academic Journal
- Accession number :
- 37844798
- Full Text :
- https://doi.org/10.1016/j.diabres.2023.110956