Back to Search Start Over

Endoscopic Ruler for varix size measurement: A multicenter pilot study.

Authors :
Huang YF
Hu SJ
Bu Y
Li YL
Deng YH
Hu JP
Yang SQ
Shen Q
McAlindon M
Shi RC
Li XQ
Song TY
Qi HL
Jiao TW
Liu MY
He F
Zhu J
Ma B
Yu XB
Guo JY
Yu YH
Yong HJ
Yao WT
Ye T
Wang H
Dong WF
Liu JG
Wei Q
Tian J
Li XG
Dray X
Qi XL
Source :
World journal of gastrointestinal endoscopy [World J Gastrointest Endosc] 2023 Sep 16; Vol. 15 (9), pp. 564-573.
Publication Year :
2023

Abstract

Background: We invented Endoscopic Ruler, a new endoscopic device to measure the size of varices in patients with cirrhosis and portal hypertension.<br />Aim: To assess the feasibility and safety of Endoscopic Ruler, and evaluate the agreement on identifying large oesophageal varices (OV) between Endoscopic Ruler and the endoscopists, as well as the interobserver agreement on diagnosing large OV using Endoscopic Ruler.<br />Methods: We prospectively and consecutively enrolled patients with cirrhosis from 11 hospitals, all of whom got esophagogastroduodenoscopy (EGD) with Endoscopic Ruler. The primary study outcome was a successful measurement of the size of varices using Endoscopic Ruler. The secondary outcomes included adverse events, operation time, the agreement of identifying large OV between the objective measurement of Endoscopic Ruler and the empirical reading of endoscopists, together with the interobserver agreement on diagnosing large OV by Endoscopic Ruler.<br />Results: From November 2020 to April 2022, a total of 120 eligible patients with cirrhosis were recruited and all of them underwent EGD examinations with Endoscopic Ruler successfully without any adverse event. The median operation time of Endoscopic Ruler was 3.00 min [interquartile range (IQR): 3.00 min]. The kappa value between Endoscopic Ruler and the endoscopists while detecting large OV was 0.52, demonstrating a moderate agreement. The kappa value for diagnosing large OV using Endoscopic Ruler among the six independent observers was 0.77, demonstrating a substantial agreement.<br />Conclusion: The data demonstrates that Endoscopic Ruler is feasible and safe for measuring the size of varices in patients with cirrhosis and portal hypertension. Endoscopic Ruler is potential to promote the clinical practice of the two-grade classification system of OV.<br />Competing Interests: Conflict-of-interest statement: All authors declared no conflicts of interest related to this study.<br /> (©The Author(s) 2023. Published by Baishideng Publishing Group Inc. All rights reserved.)

Details

Language :
English
ISSN :
1948-5190
Volume :
15
Issue :
9
Database :
MEDLINE
Journal :
World journal of gastrointestinal endoscopy
Publication Type :
Academic Journal
Accession number :
37744321
Full Text :
https://doi.org/10.4253/wjge.v15.i9.564