Back to Search
Start Over
Realization of a pulsed optically pumped Rb clock with a frequency stability below [Formula: see text].
- Source :
-
Scientific reports [Sci Rep] 2023 Aug 10; Vol. 13 (1), pp. 12974. Date of Electronic Publication: 2023 Aug 10. - Publication Year :
- 2023
-
Abstract
- We present the frequency stability performances of a vapor cell Rb clock based on the pulsed optically pumping (POP) technique. The clock has been developed in the frame of a collaboration between INRIM and Leonardo SpA, aiming to realize a space-qualified POP frequency standard. The results here reported were obtained with an engineered physics package, specifically designed for space applications, joint to laboratory-grade optics and electronics. The measured frequency stability expressed in terms of Allan deviation is [Formula: see text] at 1s and achieves the value of [Formula: see text] for integration times of 40000 s (drift removed). This is, to our knowledge, a record result for a vapor-cell frequency standard. In the paper, we show that in order to get this result, a careful stabilization of microwave and laser pulses is required.<br /> (© 2023. The Author(s).)
Details
- Language :
- English
- ISSN :
- 2045-2322
- Volume :
- 13
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Scientific reports
- Publication Type :
- Academic Journal
- Accession number :
- 37563339
- Full Text :
- https://doi.org/10.1038/s41598-023-39942-5