Comparative finite element analysis of posterior short segment fixation constructs with or without intermediate screws in the fractured vertebrae for the treatment of type a thoracolumbar fracture.
- Source :
-
Computer methods in biomechanics and biomedical engineering [Comput Methods Biomech Biomed Engin] 2024 Aug; Vol. 27 (11), pp. 1398-1409. Date of Electronic Publication: 2023 Aug 08. - Publication Year :
- 2024
-
Abstract
- Six-screw short-segment posterior fixation for thoracolumbar fractures, which involves intermediate screws at the fractured vertebrae has been proposed to reduce the rates of kyphosis recurrence and implant failure. Yet, little is known about the mechanisms and biomechanical responses by which intermediate screws at the fracture vertebrae enhance fixation strength. The objective of this study was to investigate the biomechanical properties that are associated with the augmentation of intermediate screws in relation to the severity of type A thoracolumbar fracture using finite element analysis. Short-segment stabilization models with or without augmentation screws at fractured vertebrae were established based on finite element model of moderate compressive fractures, severe compressive fractures and burst fractures. The spinal stiffness, stresses at the implanted hardware, and axial displacement of the bony defect were measured and compared under mechanical loading conditions. All six-screw stabilization showed a decreased range of motion in extension, lateral bending, and axial rotation compared to the traditional four-screw fixation models. Burst thoracolumbar fracture benefited more from augmentation of intermediate screws at the fracture vertebrae. The stress of the rod in six-screw models increased while decreased that of pedicle screws. Our results suggested that patients with more unstable fractures might achieve greater benefits from augmentation of intermediate screws at the fracture vertebrae. Augmentation of intermediate screws at the fracture vertebrae is recommended for patients with higher wedge-shaped or burst fractures to reduce the risk of hardware failure and postoperative re-collapse of injured vertebrae.
- Subjects :
- Humans
Biomechanical Phenomena
Stress, Mechanical
Finite Element Analysis
Thoracic Vertebrae surgery
Thoracic Vertebrae physiopathology
Thoracic Vertebrae injuries
Spinal Fractures surgery
Spinal Fractures physiopathology
Lumbar Vertebrae surgery
Lumbar Vertebrae physiopathology
Bone Screws
Fracture Fixation, Internal methods
Subjects
Details
- Language :
- English
- ISSN :
- 1476-8259
- Volume :
- 27
- Issue :
- 11
- Database :
- MEDLINE
- Journal :
- Computer methods in biomechanics and biomedical engineering
- Publication Type :
- Academic Journal
- Accession number :
- 37553841
- Full Text :
- https://doi.org/10.1080/10255842.2023.2243360