Back to Search Start Over

Clinical Quality Control of MRI Total Kidney Volume Measurements in Autosomal Dominant Polycystic Kidney Disease.

Authors :
Zhu C
Dev H
Sharbatdaran A
He X
Shimonov D
Chevalier JM
Blumenfeld JD
Wang Y
Teichman K
Shih G
Goel A
Prince MR
Source :
Tomography (Ann Arbor, Mich.) [Tomography] 2023 Jul 12; Vol. 9 (4), pp. 1341-1355. Date of Electronic Publication: 2023 Jul 12.
Publication Year :
2023

Abstract

Total kidney volume measured on MRI is an important biomarker for assessing the progression of autosomal dominant polycystic kidney disease and response to treatment. However, we have noticed that there can be substantial differences in the kidney volume measurements obtained from the various pulse sequences commonly included in an MRI exam. Here we examine kidney volume measurement variability among five commonly acquired MRI pulse sequences in abdominal MRI exams in 105 patients with ADPKD. Right and left kidney volumes were independently measured by three expert observers using model-assisted segmentation for axial T2, coronal T2, axial single-shot fast spin echo (SSFP), coronal SSFP, and axial 3D T1 images obtained on a single MRI from ADPKD patients. Outlier measurements were analyzed for data acquisition errors. Most of the outlier values (88%) were due to breathing during scanning causing slice misregistration with gaps or duplication of imaging slices ( n = 35), slice misregistration from using multiple breath holds during acquisition ( n = 25), composing of two overlapping acquisitions ( n = 17), or kidneys not entirely within the field of view ( n = 4). After excluding outlier measurements, the coefficient of variation among the five measurements decreased from 4.6% pre to 3.2%. Compared to the average of all sequences without errors, TKV measured on axial and coronal T2 weighted imaging were 1.2% and 1.8% greater, axial SSFP was 0.4% greater, coronal SSFP was 1.7% lower and axial T1 was 1.5% lower than the mean, indicating intrinsic measurement biases related to the different MRI contrast mechanisms. In conclusion, MRI data acquisition errors are common but can be identified using outlier analysis and excluded to improve organ volume measurement consistency. Bias toward larger volume measurements on T2 sequences and smaller volumes on axial T1 sequences can also be mitigated by averaging data from all error-free sequences acquired.

Details

Language :
English
ISSN :
2379-139X
Volume :
9
Issue :
4
Database :
MEDLINE
Journal :
Tomography (Ann Arbor, Mich.)
Publication Type :
Academic Journal
Accession number :
37489475
Full Text :
https://doi.org/10.3390/tomography9040107