Back to Search Start Over

Optimization of preparative hydrophobic interaction chromatographic purification methods.

Authors :
Gooding DL
Schmuck MN
Nowlan MP
Gooding KM
Source :
Journal of chromatography [J Chromatogr] 1986 May 30; Vol. 359, pp. 331-7.
Publication Year :
1986

Abstract

The chromatographic behavior of five proteins on hydrophobic interaction matrices having six different ligand arms was investigated using gradient elution with ammonium sulfate and ammonium acetate buffers at two pH values. The nature of the mobile phase and/or the ligand chain arm of the matrix was found to have substantial effect on the resolution, retention, and selectivity. Ovalbumin was moderately or highly retained with ammonium sulfate on all columns; however, with ammonium acetate, ovalbumin was not retained on SynChropak Hydroxypropyl and Propyl columns. Chromatographic conditions developed for analytical hydrophobic interaction chromatography columns containing 6.5-micron packings were adapted to preparative columns packed with 30-micron SynChroprep packings for the separation of serum components. Dynamic load capacities were 4-13 mg of ovalbumin per ml of column volume.

Details

Language :
English
Volume :
359
Database :
MEDLINE
Journal :
Journal of chromatography
Publication Type :
Academic Journal
Accession number :
3733935
Full Text :
https://doi.org/10.1016/0021-9673(86)80087-5