Quality control of azure B preparations by liquid chromatography and standardization with azure B tetrafluoroborate.
- Source :
-
Clinica chimica acta; international journal of clinical chemistry [Clin Chim Acta] 1986 May 15; Vol. 156 (3), pp. 247-57. - Publication Year :
- 1986
-
Abstract
- A liquid chromatographic procedure for the quantitative determination of the thiazine dye azure B, the principal constituent of Romanowsky stains, is presented. Unlike previous methods relying on peak area normalization, the present approach involves real quantitation through calibration with the reference standard azure B tetrafluoroborate. The method has been used for the quality control of commercial azure B preparations and to study their stability in stock and staining solutions, either or not in the presence of eosin Y. Results suggest that highly pure azure B perchlorate meets the requirements of a reference material, useful for standardization of Romanowsky-Giemsa staining in haematology.
- Subjects :
- Azure Stains analysis
Calibration
Chromatography, High Pressure Liquid
Enkephalin, Methionine administration & dosage
Enkephalin, Methionine pharmacology
Quality Control
Reference Standards
Solutions
Azure Stains standards
Enkephalin, Methionine analogs & derivatives
Phenothiazines standards
Subjects
Details
- Language :
- English
- ISSN :
- 0009-8981
- Volume :
- 156
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Clinica chimica acta; international journal of clinical chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 3719983
- Full Text :
- https://doi.org/10.1016/0009-8981(86)90068-9