Simple and effective immobilization for radiation treatment of choroidal melanoma.

Authors :
van Prooijen M
Chau L
Tsui G
Kelly V
Holwell M
Tadic T
Tsang DS
Krema H
Laperriere N
Source :
Medical dosimetry : official journal of the American Association of Medical Dosimetrists [Med Dosim] 2023 Autumn; Vol. 48 (3), pp. 154-160. Date of Electronic Publication: 2023 Apr 27.
Publication Year :
2023

Abstract

At our institution, patients diagnosed with choroidal melanoma requiring external beam radiation therapy are treated with two 6 MV volumetric-modulated arcs delivering 50 Gy over 5 daily fractions. The patient is immobilized using an Orfit head and neck mask and is directed to look at a light emitting diode (LED) during CT simulation and treatment to minimize eye movement. Patient positioning is checked with cone beam computed tomography (CBCT) daily. Translational and rotational displacements greater than 1 mm or 1° off the planned isocenter position are corrected using a Hexapod couch. The aim of this study is to verify that the mask system provides adequate immobilization and to verify our 2-mm planning target volume (PTV) margins are sufficient. Residual displacements provided by pretreatment verification and post-treatment CBCT data sets were used to assess the impact of patient mobility during treatment on the reconstructed delivered dose to the target and organs at risk. The PTV margin calculated using van Herk's method <superscript>1</superscript> was used to assess patient motion plus other factors that affect treatment position, such as kV-MV isocenter coincidence. Patient position variations were small and were shown to not cause significant dose variations between the planned and reconstructed dose to the target and organs at risk. The PTV margin analysis showed patient translational motion alone required a PTV margin of 1 mm. Given other factors that affect treatment delivery accuracy, a 2-mm PTV margin was shown to be sufficient for treatment of 95% of our patients with 100% of dose delivered to the GTV. The mask immobilization with LED focus is robust and we showed a 2-mm PTV margin is adequate with this technique.<br />Competing Interests: Conflicts of Interest The authors declare no conflicts of interest.<br /> (Copyright © 2023 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1873-4022
Volume :
48
Issue :
3
Database :
MEDLINE
Journal :
Medical dosimetry : official journal of the American Association of Medical Dosimetrists
Publication Type :
Academic Journal
Accession number :
37120386
Full Text :
https://doi.org/10.1016/j.meddos.2023.03.002