Clinical Linear Accelerator-Based Electron FLASH: Pathway for Practical Translation to FLASH Clinical Trials.

Authors :
No HJ
Wu YF
Dworkin ML
Manjappa R
Skinner L
Ashraf MR
Lau B
Melemenidis S
Viswanathan V
Yu AS
Surucu M
Schüler E
Graves EE
Maxim PG
Loo BW Jr
Source :
International journal of radiation oncology, biology, physics [Int J Radiat Oncol Biol Phys] 2023 Oct 01; Vol. 117 (2), pp. 482-492. Date of Electronic Publication: 2023 Apr 25.
Publication Year :
2023

Abstract

Purpose: Ultrahigh-dose-rate (UHDR) radiation therapy (RT) has produced the FLASH effect in preclinical models: reduced toxicity with comparable tumor control compared with conventional-dose-rate RT. Early clinical trials focused on UHDR RT feasibility using specialized devices. We explore the technical feasibility of practical electron UHDR RT on a standard clinical linear accelerator (LINAC).<br />Methods and Materials: We tuned the program board of a decommissioned electron energy for UHDR electron delivery on a clinical LINAC without hardware modification. Pulse delivery was controlled using the respiratory gating interface. A short source-to-surface distance (SSD) electron setup with a standard scattering foil was configured and tested on an anthropomorphic phantom using circular blocks with 3- to 20-cm field sizes. Dosimetry was evaluated using radiochromic film and an ion chamber profiler.<br />Results: UHDR open-field mean dose rates at 100, 80, 70, and 59 cm SSD were 36.82, 59.52, 82.01, and 112.83 Gy/s, respectively. At 80 cm SSD, mean dose rate was ∼60 Gy/s for all collimated field sizes, with an R80 depth of 6.1 cm corresponding to an energy of 17.5 MeV. Heterogeneity was <5.0% with asymmetry of 2.2% to 6.2%. The short SSD setup was feasible under realistic treatment conditions simulating broad clinical indications on an anthropomorphic phantom.<br />Conclusions: Short SSD and tuning for high electron beam current on a standard clinical LINAC can deliver flat, homogenous UHDR electrons over a broad, clinically relevant range of field sizes and depths with practical working distances in a configuration easily reversible to standard clinical use.<br /> (Copyright © 2023 Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1879-355X
Volume :
117
Issue :
2
Database :
MEDLINE
Journal :
International journal of radiation oncology, biology, physics
Publication Type :
Academic Journal
Accession number :
37105403
Full Text :
https://doi.org/10.1016/j.ijrobp.2023.04.011