Back to Search Start Over

Quantitative longitudinal mapping of radiation-treated prostate cancer using MR fingerprinting with radial acquisition and subspace reconstruction.

Authors :
Yu VY
Otazo R
Wu C
Subashi E
Baumann M
Koken P
Doneva M
Mazurkewitz P
Shasha D
Zelefsky M
Cervino L
Cohen O
Source :
Magnetic resonance imaging [Magn Reson Imaging] 2023 Sep; Vol. 101, pp. 25-34. Date of Electronic Publication: 2023 Apr 02.
Publication Year :
2023

Abstract

MR fingerprinting (MRF) enables fast multiparametric quantitative imaging with a single acquisition and has been shown to improve diagnosis of prostate cancer. However, most prostate MRF studies were performed with spiral acquisitions that are sensitive to B0 inhomogeneities and consequent blurring. In this work, a radial MRF acquisition with a novel subspace reconstruction technique was developed to enable fast T1/T2 mapping in the prostate in under 4 min. The subspace reconstruction exploits the extensive temporal correlations in the MRF dictionary to pre-compute a low dimensional space for the solution and thus reduce the number of radial spokes to accelerate the acquisition. Iterative reconstruction with the subspace model and additional regularization of the signal representation in the subspace is performed to minimize the number of spokes and maintain matching quality and SNR. Reconstruction accuracy was assessed using the ISMRM NIST phantom. In-vivo validation was performed on two healthy subjects and two prostate cancer patients undergoing radiation therapy. The longitudinal repeatability was quantified using the concordance correlation coefficient (CCC) in one of the healthy subjects by repeated scans over 1 year. One prostate cancer patient was scanned at three time points, before initiating therapy and following brachytherapy and external beam radiation. Changes in the T1/T2 maps obtained with the proposed method were quantified. The prostate, peripheral and transitional zones, and visible dominant lesion were delineated for each study, and the statistics and distribution of the quantitative mapping values were analyzed. Significant image quality improvements compared with standard reconstruction methods were obtained with the proposed subspace reconstruction method. A notable decrease in the spread of the T1/T2 values without biasing the estimated mean values was observed with the subspace reconstruction and agreed with reported literature values. The subspace reconstruction enabled visualization of small differences in T1/T2 values in the tumor region within the peripheral zone. Longitudinal imaging of a volunteer subject yielded CCC of 0.89 for MRF T1, and 0.81 for MRF T2 in the prostate gland. Longitudinal imaging of the prostate patient confirmed the feasibility of capturing radiation treatment related changes. This work is a proof-of-concept for a high resolution and fast quantitative mapping using golden-angle radial MRF combined with a subspace reconstruction technique for longitudinal treatment response assessment in subjects undergoing radiation treatment.<br />Competing Interests: Declaration of Competing Interest This work was performed as part of a master research agreement between MSKCC and Philips Healthcare. MB, MD, PM, and PK were employees of Philips Healthcare while making contributions to this work.<br /> (Copyright © 2023 Elsevier Inc. All rights reserved.)

Details

Language :
English
ISSN :
1873-5894
Volume :
101
Database :
MEDLINE
Journal :
Magnetic resonance imaging
Publication Type :
Academic Journal
Accession number :
37015305
Full Text :
https://doi.org/10.1016/j.mri.2023.03.019