Cerebrovascular reactivity measurements using 3T BOLD MRI and a fixed inhaled CO 2 gas challenge: Repeatability and impact of processing strategy.

Authors :
Sleight E
Stringer MS
Mitchell I
Murphy M
Marshall I
Wardlaw JM
Thrippleton MJ
Source :
Frontiers in physiology [Front Physiol] 2023 Feb 06; Vol. 14, pp. 1070233. Date of Electronic Publication: 2023 Feb 06 (Print Publication: 2023).
Publication Year :
2023

Abstract

Introduction: Cerebrovascular reactivity (CVR) measurements using blood oxygen level dependent (BOLD) magnetic resonance imaging (MRI) are commonly used to assess the health of cerebral blood vessels, including in patients with cerebrovascular diseases; however, evidence and consensus regarding reliability and optimal processing are lacking. We aimed to assess the repeatability, accuracy and precision of voxel- and region-based CVR measurements at 3 T using a fixed inhaled (FI) CO <subscript>2</subscript> stimulus in a healthy cohort. Methods: We simulated the effect of noise, delay constraints and voxel- versus region-based analysis on CVR parameters. Results were verified in 15 healthy volunteers (28.1±5.5 years, female: 53%) with a test-retest MRI experiment consisting of two CVR scans. CVR magnitude and delay in grey matter (GM) and white matter were computed for both analyses assuming a linear relationship between the BOLD signal and time-shifted end-tidal CO <subscript>2</subscript> (EtCO <subscript>2</subscript> ) profile. Results: Test-retest repeatability was high [mean (95% CI) inter-scan difference: -0.01 (-0.03, -0.00) %/mmHg for GM CVR magnitude; -0.3 (-1.2,0.6) s for GM CVR delay], but we detected a small systematic reduction in CVR magnitude at scan 2 versus scan 1, accompanied by a greater EtCO2 change [±1.0 (0.4,1.5) mmHg] and lower heart rate [-5.5 (-8.6,-2.4] bpm]. CVR magnitude estimates were higher for voxel- versus region-based analysis [difference in GM: ±0.02 (0.01,0.03) %/mmHg]. Findings were supported by simulation results, predicting a positive bias for voxel-based CVR estimates dependent on temporal contrast-to-noise ratio and delay fitting constraints and an underestimation for region-based CVR estimates. Discussion: BOLD CVR measurements using FI stimulus have good within-day repeatability in healthy volunteers. However, measurements may be influenced by physiological effects and the analysis protocol. Voxel-based analyses should be undertaken with care due to potential for systematic bias; region-based analyses are more reliable in such cases.<br />Competing Interests: The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.<br /> (Copyright © 2023 Sleight, Stringer, Mitchell, Murphy, Marshall, Wardlaw and Thrippleton.)

Details

Language :
English
ISSN :
1664-042X
Volume :
14
Database :
MEDLINE
Journal :
Frontiers in physiology
Publication Type :
Academic Journal
Accession number :
36814481
Full Text :
https://doi.org/10.3389/fphys.2023.1070233