A stepwise external cardioversion protocol for atrial fibrillation to maximize acute success rate.

Authors :
Darrat Y
Leung S
Elayi L
Parrott K
Ogunbayo G
Kotter J
Sorrell V
Gupta V
Anaya P
Morales G
Catanzarro J
Delisle B
Elayi CS
Source :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2023 Mar 30; Vol. 25 (3), pp. 828-834.
Publication Year :
2023

Abstract

Aims: Cardioversion is a very commonly performed procedure for persistent atrial fibrillation (AF). However, there is no well-defined protocol to address failed external electrical direct current cardioversion. The aim of the study is to test the efficacy of a pre-defined stepwise cardioversion protocol for patients with persistent AF of ≤12 months. Success was the achievement of sinus rhythm.<br />Methods and Results: The study population included patients with persistent AF of ≤12 months duration requiring rhythm management. Patients were offered cardioversion using a pre-defined stepwise protocol using different electrode placement locations, applying compression at end of expiration, and higher energy delivered simultaneously through two defibrillators. : A total of 414 patients were included in the study, of which 362 (87.4%) required a single successful cardioversion. The remaining 52 (12.5%) patients required additional cardioversion attempts using the stepwise cardioversion protocol with an overall success rate of 99.3%. Two simultaneous defibrillators were required in 14 patients (3.4%). Patients with multiple cardioversions (13.5%) experienced more local skin irritation and pain compared with patients with single cardioversion (13.5% vs. 3.5%, P = 0.004). The predictor for the need for multiple cardioversion attempts is high body mass index, while high transthoracic impedance is associated with failed cardioversion. No major complications were observed during the study.<br />Conclusion: The stepwise cardioversion protocol has a high success rate of >99% and can be safely performed in outpatient or inpatient settings.<br />Competing Interests: Conflict of interest: All authors declare that they have no conflict of interest.<br /> (© The Author(s) 2023. Published by Oxford University Press on behalf of the European Society of Cardiology.)

Details

Language :
English
ISSN :
1532-2092
Volume :
25
Issue :
3
Database :
MEDLINE
Journal :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
36748366
Full Text :
https://doi.org/10.1093/europace/euad009