Validation of a novel point-of-care test for alanine aminotransferase measurement: A pilot cohort study.

Authors :
Howell J
Van H
Pham MD
Sawhney R
Li F
Bhat P
Lubel J
Kemp W
Bloom S
Majumdar A
McCaughan GW
Hall S
Spelman T
Doyle JS
Hellard M
Visvanathan K
Thompson A
Drummer HE
Anderson D
Source :
Liver international : official journal of the International Association for the Study of the Liver [Liver Int] 2023 May; Vol. 43 (5), pp. 989-999. Date of Electronic Publication: 2023 Feb 16.
Publication Year :
2023

Abstract

Background: Alanine aminotransferase (ALT) measurement is essential for evaluation of liver disease. We validated a novel rapid point-of-care (POC) test for ALT1 against laboratory ALT.<br />Methods: Stored plasma samples from adults with chronic liver disease (Test cohort n = 240; Validation cohort n = 491) were analysed using the BioPoint® antigen immunoassay POC ALT1 lateral flow test, which provides quantitative ALT results (Axxin handheld reader) or semi-quantitative results (visual read, cut off 40 IU/ml). The accuracy of POC ALT1 to detect ALT > 40 IU/L was determined by ROC analysis. In patients with chronic hepatitis B, treatment eligibility (EASL criteria) was determined using POC ALT1 and compared to laboratory ALT.<br />Results: POC ALT1 test had good accuracy for laboratory ALT > 40 IU/L: AUROC 0.93 (95% CI: 0.89-0.96) in the Test cohort and AUROC 0.92 (95% CI: 0.88-0.95) in the Validation cohort. POC ALT1 cut off of 0.8 for ALT > 40 IU/L maximised sensitivity (97%) and specificity (71%) in the Test cohort (42% laboratory ALT > 40 IU/L) and yielded PPV 84% and NPV 91% in the Validation cohort (19% laboratory ALT > 40 IU/L). Semi-quantitative POC ALT1 had good accuracy for laboratory ALT in the Validation cohort (AUROC 0.85, 95% CI: 0.81-0.99; sensitivity 77% and specificity 93%). Combined with HBV DNA and transient elastography, both quantitative and semi-quantitative POC ALT1 tests had good accuracy for excluding hepatitis B treatment needs (sensitivity 96%, specificity 78% and NPV 99%).<br />Conclusion: The POC ALT1 test had good accuracy for elevated ALT levels and for determining treatment eligibility among people with chronic hepatitis B.<br /> (© 2023 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.)

Details

Language :
English
ISSN :
1478-3231
Volume :
43
Issue :
5
Database :
MEDLINE
Journal :
Liver international : official journal of the International Association for the Study of the Liver
Publication Type :
Academic Journal
Accession number :
36719055
Full Text :
https://doi.org/10.1111/liv.15531