Back to Search
Start Over
Multicentric study on surgical information and early safety and performance results with the Bonebridge BCI 602: an active transcutaneous bone conduction hearing implant.
- Source :
-
European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery [Eur Arch Otorhinolaryngol] 2023 Apr; Vol. 280 (4), pp. 1565-1579. Date of Electronic Publication: 2023 Jan 10. - Publication Year :
- 2023
-
Abstract
- Aim: This European multicentric study aimed to prove safety and performance of the Bonebridge BCI 602 in children and adults suffering from either conductive hearing loss (CHL), mixed hearing loss (MHL), or single-sided sensorineural deafness (SSD).<br />Methods: 33 patients (13 adults and 10 children with either CHL or MHL and 10 patients with SSD) in three study groups were included. Patients were their own controls (single-subject repeated measures), comparing the unaided or pre-operative to the 3-month post-operative outcomes. Performance was evaluated by sound field thresholds (SF), word recognition scores (WRS) and/or speech reception thresholds in quiet (SRT) and in noise (SNR). Safety was demonstrated with a device-specific surgical questionnaire, adverse event reporting and stable pure-tone measurements.<br />Results: The Bonebridge BCI 602 significantly improved SF thresholds (+ 25.5 dB CHL/MHL/SSD), speech intelligibility in WRS (+ 68.0% CHL/MHL) and SRT in quiet (- 16.5 dB C/MHL) and in noise (- 3.51 dB SNR SSD). Air conduction (AC) and bone conduction (BC) thresholds remained stable over time. All adverse events were resolved, with none unanticipated. Mean audio processor wearing times in hours [h] per day for the CHL/MHL group were ~ 13 h for adults, ~ 11 h for paediatrics and ~ 6 h for the SSD group. The average surgical length was 57 min for the CHL/MHL group and 42 min for the SSD group. The versatility of the BCI 602 (reduced drilling depth and ability to bend the transition for optimal placement) allows for treatment of normal, pre-operated and malformed anatomies. All audiological endpoints were reached.<br />Conclusions: The Bonebridge BCI 602 significantly improved hearing thresholds and speech understanding. Since implant placement follows the patient's anatomy instead of the shape of the device and the duration of surgery is shorter than with its predecessor, implantation is easier with the BCI 602. Performance and safety were proven for adults and children as well as for the CHL/MHL and SSD indications 3 months post-operatively.<br /> (© 2023. The Author(s).)
- Subjects :
- Adult
Humans
Child
Bone Conduction
Hearing
Hearing Loss, Conductive surgery
Treatment Outcome
Multicenter Studies as Topic
Brain-Computer Interfaces
Hearing Aids
Hearing Loss, Mixed Conductive-Sensorineural surgery
Deafness surgery
Hearing Loss surgery
Hearing Loss, Sensorineural surgery
Speech Perception
Subjects
Details
- Language :
- English
- ISSN :
- 1434-4726
- Volume :
- 280
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery
- Publication Type :
- Academic Journal
- Accession number :
- 36625869
- Full Text :
- https://doi.org/10.1007/s00405-022-07792-y