Development of a three-dimensional dose evaluation method for computed tomography.

Authors :
Katsuda T
Gotanda R
Gotanda T
Kuwano T
Tanki N
Yabunaka K
Source :
Journal of applied clinical medical physics [J Appl Clin Med Phys] 2023 Feb; Vol. 24 (2), pp. e13897. Date of Electronic Publication: 2023 Jan 07.
Publication Year :
2023

Abstract

During a single scan using computed tomography, an X-ray tube orbits along a 360°-circular path around the patient. A scan obtained using the half-cylindrical type phantoms with a radiochromic film sandwiched in between reveals a pixel value map illustrating the two-dimensional (2D) dose distribution. A three-dimensional (3D) dose distribution can be obtained with a 360° rotation of the 2D dose map. This study evaluates the concept and methodology of creating a 3D dose map to develop a phantom with a radiochromic film for obtaining the 3D dose distribution. The coronal and axial plane dose distributions were also evaluated. A single scan computed tomography image obtained using a half-cylindrical type of acrylic phantom with a sandwiched radiochromic film was studied. The diameters of the phantoms were 10 and 16 cm, and their lengths were 30 cm. A 2D image of the XR-QA2 film was obtained using an image scanner and image processing software. A red channel image was used to obtain the 3D dose distribution using a computing platform. A pseudo color was applied to the red channel image from which cross-sectional color images were obtained. Half of the cross-sectional pixel data were rotated by 360° to generate the data for each axial plane. The image created was saved, and a 3D pixel value map was constructed. The dose measurement procedure for the 3D dose distribution was developed using half-cylindrical acrylic phantoms with a radiochromic film.<br /> (© 2023 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.)

Details

Language :
English
ISSN :
1526-9914
Volume :
24
Issue :
2
Database :
MEDLINE
Journal :
Journal of applied clinical medical physics
Publication Type :
Academic Journal
Accession number :
36610053
Full Text :
https://doi.org/10.1002/acm2.13897