COMPARISON OF THE 64- AND 80-DETECTOR ROW COMPUTED TOMOGRAPHY AMONG THE CT NUMBER AND RADIATION DOSE DURING LOWER EXTREMITY COMPUTED TOMOGRAPHY ANGIOGRAPHY.
- Source :
-
Radiation protection dosimetry [Radiat Prot Dosimetry] 2023 Mar 17; Vol. 199 (4), pp. 356-362. - Publication Year :
- 2023
-
Abstract
- To compare the computed tomography (CT) number and the radiation dose between the 64 (group A) and 80-detector row (group B) during lower extremity computed tomography angiography (LE-CTA). We enrolled 144 patients underwent LE-CTA and compared the CT number for the popliteal arteries, radiation dose and the rate of the optimal CT number during the LE-CTA exceeding 200 HU between the two groups. The CT number for the popliteal arteries and mean dose-length product was significantly higher in Group A than in Group B (P < 0.01). The rate of the optimal CT number for the popliteal arteries was 23.6% with Group B scanner and 56.9% with Group A (P < 0.05). The 64-detector row CT was significantly higher in the CT number for the popliteal arteries, radiation dose and rate of the optimal CT number during the LE-CTA than the 80-detector row. Depiction ability did not improve by using a high CT scanner with a wider detector during LE-CTA.<br /> (© The Author(s) 2022. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)
Details
- Language :
- English
- ISSN :
- 1742-3406
- Volume :
- 199
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Radiation protection dosimetry
- Publication Type :
- Academic Journal
- Accession number :
- 36609697
- Full Text :
- https://doi.org/10.1093/rpd/ncac277