Back to Search
Start Over
Risk factors for in-stent restenosis in maintenance hemodialysis patients with central venous occlusive disease and biomechanical assessment of stents.
- Source :
-
The journal of vascular access [J Vasc Access] 2024 May; Vol. 25 (3), pp. 943-952. Date of Electronic Publication: 2022 Dec 20. - Publication Year :
- 2024
-
Abstract
- Objectives: To investigate the risk factors and biomechanical mechanisms of in-stent restenosis (ISR) in central venous occlusive disease (CVOD).<br />Patients and Methods: This retrospective study consecutively included 77 maintenance hemodialysis (MHD) patients with CVOD who received the first percutaneous transluminal angioplasty with stenting (PTS) due to symptomatic CVOD in a tertiary hospital. The mean age was 59.7 ± 14.0 years, and 51.9% of patients were male. The clinical characteristics, occurrence of ISR and patency rates were recorded. Finite element method was applied to assess the biomechanical properties of stents.<br />Results: Among 77 patients with a mean CVS score of 8.0 ± 2.8, 20.8%, 62.3%, and 16.9% of patients had the main vein of CVOD in the subclavian vein, brachiocephalic vein, and superior vena cava, respectively. A total of 72 (93.5%) patients received successful PTS treatment, for which the stents implanted were mainly Fluency covered stent (48.6%) and SMART bare stent (31.9%). During 15 (10-24)-months of follow-up, ISR occurred in 36.1% of the 72 patients. The primary and assisted primary patency rates at 6, 12, and 18 months were 78%, 56%, 42% and 95%, 90%, 87%, respectively. A prolonged dialysis vintage was an independent risk factor for ISR, yet the stent type or the main vein location was not correlated with ISR. Among three laser-engraving stents, the SMART stent was the best in terms of flexibility, stress, and strain on stents but worst in stress or strain on vessels. The Luminexx stent was the best in radial force and worst in stress or strain on stents. The Vici stent was the best in stress and strain on vessels and worst in radial force and flexibility.<br />Conclusions: An unsatisfactory comprehensive biomechanical performance from configurations rooted in existing stents may account for the high incidence of ISR in CVOD.<br />Competing Interests: Declaration of conflicting interestsThe author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
- Subjects :
- Humans
Male
Middle Aged
Risk Factors
Female
Retrospective Studies
Aged
Treatment Outcome
Biomechanical Phenomena
Time Factors
Finite Element Analysis
Constriction, Pathologic
Models, Cardiovascular
Vascular Diseases physiopathology
Vascular Diseases etiology
Vascular Diseases therapy
Angioplasty, Balloon instrumentation
Angioplasty, Balloon adverse effects
Risk Assessment
Adult
Patient-Specific Modeling
Renal Dialysis
Stents
Vascular Patency
Prosthesis Design
Recurrence
Subjects
Details
- Language :
- English
- ISSN :
- 1724-6032
- Volume :
- 25
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- The journal of vascular access
- Publication Type :
- Academic Journal
- Accession number :
- 36540050
- Full Text :
- https://doi.org/10.1177/11297298221139640