Back to Search Start Over

Evaluation of interference from 16 hemoglobin variants on hemoglobin A 1c measurement by five methods.

Authors :
Yang X
Zeng X
Zhang Y
Kuang W
He D
Source :
Scandinavian journal of clinical and laboratory investigation [Scand J Clin Lab Invest] 2023 Feb; Vol. 83 (1), pp. 18-22. Date of Electronic Publication: 2022 Dec 19.
Publication Year :
2023

Abstract

Hb variants prevalent in China are different from those in other countries. We aimed to assess the interference from Hb variants found in China on HbA1c measurement. All Hb variants were confirmed using Sanger sequencing. HbA1c was measured using a capillary electrophoresis method (Capillarys 3 OCTA), two cation-exchange high-performance liquid chromatography methods (ADAMS HA-8180V and HLC-723 G8 standard mode), an immunoassay (Cobas c501), and a boronate affinity chromatography method (Premier Hb9210). Premier Hb9210 was used as a comparative method. A total of 16 species of Hb variants were identified in 102 variant carriers. The most common variant was Hb E, followed by Hb Q-Thailand, Hb New York and Hb J-Bangkok. Clinically significant interference was observed for the Capillarys 3 OCTA (two Hb variants), ADAMS HA-8180V (seven Hb variants), HLC-723 G8 (14 Hb variants), and Cobas c501 (two Hb variants). The proportion of unacceptable HbA1c results was 13.7% for Capillarys 3 OCTA, 52.9% for HA-8180V, 83.3% for HLC-723 G8, and 3.9% for Cobas c501. Hb variants in China severely affect the accuracy of some commonly used HbA1c methods.

Details

Language :
English
ISSN :
1502-7686
Volume :
83
Issue :
1
Database :
MEDLINE
Journal :
Scandinavian journal of clinical and laboratory investigation
Publication Type :
Academic Journal
Accession number :
36534489
Full Text :
https://doi.org/10.1080/00365513.2022.2155990