Back to Search Start Over

Relapsing low-flow alarms due to suboptimal alignment of the left ventricular assist device inflow cannula.

Authors :
Zijderhand CF
Knol WG
Budde RPJ
van der Heiden CW
Veen KM
Sjatskig J
Manintveld OC
Constantinescu AA
Birim O
Bekkers JA
Bogers AJJC
Caliskan K
Source :
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery [Eur J Cardiothorac Surg] 2022 Sep 02; Vol. 62 (4).
Publication Year :
2022

Abstract

Objectives: This retrospective study investigated the correlation between the angular position of the left ventricular assist device (LVAD) inflow cannula and relapsing low-flow alarms.<br />Methods: Medical charts were reviewed of all patients with HeartMate 3 LVAD support for relapsing low-flow alarms. A standardized protocol was created to measure the angular position with a contrast-enhanced computed tomography scan. Statistics were done using a gamma frailty model with a constant rate function.<br />Results: For this analysis, 48 LVAD-supported patients were included. The majority of the patients were male (79%) with a median age of 57 years and a median follow-up of 30 months (interquartile range: 19-41). Low-flow alarm(s) were experienced in 30 (63%) patients. Angulation towards the septal-lateral plane showed a significant increase in low-flow alarms over time with a constant rate function of 0.031 increase in low-flow alarms per month of follow-up per increasing degree of angulation (P = 0.048). When dividing this group using an optimal cut-off point, a significant increase in low-flow alarms was observed when the septal-lateral angulation was 28° or more (P = 0.001). Anterior-posterior and maximal inflow cannula angulation did not show a significant difference.<br />Conclusions: This study showed an increasing number of low-flow alarms when the degrees of LVAD inflow cannula expand towards the septal-lateral plane. This emphasizes the importance of the LVAD inflow cannula angular position to prevent relapsing low-flow alarms with the risk of diminished quality of life and morbidity.<br /> (© The Author(s) 2022. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.)

Details

Language :
English
ISSN :
1873-734X
Volume :
62
Issue :
4
Database :
MEDLINE
Journal :
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
Publication Type :
Academic Journal
Accession number :
35993906
Full Text :
https://doi.org/10.1093/ejcts/ezac415