Back to Search
Start Over
Detection of interference by cyclobenzaprine in liquid-chromatographic assays of tricyclic antidepressants.
- Source :
-
Clinical chemistry [Clin Chem] 1987 Jun; Vol. 33 (6), pp. 819-20. - Publication Year :
- 1987
-
Abstract
- We evaluated a technique for detecting cyclobenzaprine interference with liquid-chromatographic assays for tricyclic antidepressants. The technique involves dual-wavelength absorbance monitoring of the column effluent at 214 and 254 nm. Ratios of analyte peak heights at each wavelength are used to check for the presence of co-eluting interferences. With this technique, one can detect interference with an amitriptyline assay caused by 10 micrograms of cyclobenzaprine per liter.
Details