Back to Search
Start Over
The First FDA Approved Early Feasibility Study of a Novel Percutaneous Bone Anchored Prosthesis for Transfemoral Amputees: A Prospective 1-year Follow-up Cohort Study.
- Source :
-
Archives of physical medicine and rehabilitation [Arch Phys Med Rehabil] 2022 Nov; Vol. 103 (11), pp. 2092-2104. Date of Electronic Publication: 2022 Jul 29. - Publication Year :
- 2022
-
Abstract
- Objective: Evaluate the safety and efficacy of a novel press-fit bone-anchored prosthesis in an FDA approved study.<br />Design: Single-center, prospective 1-year follow-up cohort study of a percutaneous bone anchored prosthesis.<br />Setting: Veterans Health Administration Hospital.<br />Participants: Ten male Veterans with unilateral transfemoral amputation that occurred at least 6 months prior to enrollment and was not the result of dysvascular disease (N=10).<br />Interventions: All participants received the novel press-fit Percutaneous Osseointegrated Prosthesis (POP) and a minimum of 10 days supervised rehabilitation therapy.<br />Outcome Measures: Adverse events and radiographs were collected to assess device safety. Temporal assessments of bone density, stomal skin, prosthetic don/doff, functional ambulation, and patient-reported outcome compared our POP to a socket suspension system.<br />Results: Ten male participants mean age 48.8±12.1 years (range, 32-68 y) with mean time since amputation of 9.4± years (range 1-18 y) completed a two-staged implantation protocol and progressed to ambulation with an assistive device by post-operative day 14. Eight of 10 completed all study procedures. One implant loosened at 5 weeks, requiring removal. A second was removed after periprosthetic fracture from a non-device-related fall at 28 weeks. One patient required oral antibiotics for superficial infection. There were significant (P<.05) increases in bone density in the lumbar spine and adjacent to the distal porous coating with no radiographic evidence of bone resorption. Compared to the socket system, the use of POP significantly (P<.05) reduced prosthetic don and doff times and patient-reported prosthetic problem scores. Significant improvements (P<.05) in mean mobility, global health, and walking test scores were also observed.<br />Conclusions: Improvements in bone density, function, and patient reported outcomes were observed with the POP device when compared to a socket suspension system. This Early Feasibility Study established initial safety and effectiveness of the POP device, supporting expanded investigation as an alternative to socket prostheses.<br /> (Published by Elsevier Inc.)
- Subjects :
- Male
Humans
Infant
Follow-Up Studies
Prospective Studies
Feasibility Studies
Osseointegration
Amputation, Surgical rehabilitation
Prosthesis Implantation adverse effects
Prosthesis Implantation methods
Prosthesis Design
Amputees rehabilitation
Bone-Anchored Prosthesis adverse effects
Artificial Limbs adverse effects
Subjects
Details
- Language :
- English
- ISSN :
- 1532-821X
- Volume :
- 103
- Issue :
- 11
- Database :
- MEDLINE
- Journal :
- Archives of physical medicine and rehabilitation
- Publication Type :
- Academic Journal
- Accession number :
- 35908658
- Full Text :
- https://doi.org/10.1016/j.apmr.2022.06.008