Back to Search Start Over

Amperometric Sensor for Selective On-Site Analysis of Free Sulfite in Wines.

Authors :
Gajdár J
Herzog G
Etienne M
Source :
ACS sensors [ACS Sens] 2022 Aug 26; Vol. 7 (8), pp. 2209-2217. Date of Electronic Publication: 2022 Jul 15.
Publication Year :
2022

Abstract

Accurate and rapid on-site analysis of free SO <subscript>2</subscript> content is crucial in the process of winemaking from a producer and consumer perspective. Herein, we present an amperometric sensor based on commercially available screen-printed electrodes coupled with an electrochemical oxygen filter. The developed amperometric method gave a linear response in a concentration range up to 200 mg L <superscript>-1</superscript> with a limit of quantification of 7.5 mg L <superscript>-1</superscript> . The applicability of the developed sensor was successfully tested on 27 white and red wine samples and compared to the Ripper method (iodometry) that is a standard procedure for free SO <subscript>2</subscript> determination. The sensor exhibits similar precision and accuracy but shows no interference from oxidizable species such as ascorbic acid, which is a major advantage over iodometric titration. The performance of the sensor was in addition positively evaluated during on-site analysis in a winery.

Details

Language :
English
ISSN :
2379-3694
Volume :
7
Issue :
8
Database :
MEDLINE
Journal :
ACS sensors
Publication Type :
Academic Journal
Accession number :
35838550
Full Text :
https://doi.org/10.1021/acssensors.2c00611