Back to Search Start Over

[Attempt to adapt measurement of serum iron levels to the Centrifichem with a direct method using bathophenanthroline sulfonate-guanidine. Value and limits].

Authors :
Plomteux G
Vernet-Nyssen M
Charlier C
Paris M
Source :
Annales de biologie clinique [Ann Biol Clin (Paris)] 1987; Vol. 45 (1), pp. 52-6.
Publication Year :
1987

Abstract

The authors have studied a direct automated colorimetric determination of iron serum on Centrifichem. Accuracy and linearity are the same than these obtained by the SFBC recommended method. Precision is poor: the coefficient of variation is about 15% for low values. No alteration in the absorbance is recognized by bilirubin, even in the serum which contains 342 mumol.l-1 of bilirubin. The value of serum iron increases significantly as hemoglobin and copper are added to the serum.