Back to Search
Start Over
The Performances of Conventional Titanium and Silver-Coated Megaprostheses in Non-oncological and Post-oncological Patients: An Analysis of Infection Failures in 142 Patients.
- Source :
-
The archives of bone and joint surgery [Arch Bone Jt Surg] 2022 May; Vol. 10 (5), pp. 439-446. - Publication Year :
- 2022
-
Abstract
- Background: Megaprostheses are one of the preferred choices of reconstruction after tumor resection. Periprosthetic joint infections are one of the most serious complications of joint prostheses surgeries. In this study, our aim was to analyze the efficacy of silver-coated megaprostheses in reducing the risk of prosthesis-related infection.<br />Methods: One hundred forty-two patients who had undergone implantation of a mega-endoprosthesis for non-neoplastic or post-neoplastic conditions were included in this retrospective study. The end-point of the survival analysis was the prosthesis failure due to infection.<br />Results: Thirty-eight patients had undergone implantation of a silver-coated megaprosthesis and 104 patients a megaprosthesis without silver coating. The survival analysis showed an overall infection-free survival rate of 82.3% at five years and 61.9% at 10 years. Silver-coated prostheses had an HR of 0.72 (95% CI: 0.26-2.05; P=0.54 ).<br />Conclusion: Implantation of a silver-coated mega-prosthesis in non-oncological patients did not significantly reduce the risk of prosthesis-related infection.<br />Competing Interests: All authors declare that they have no conflict of interest related to the publication of this paper.
Details
- Language :
- English
- ISSN :
- 2345-4644
- Volume :
- 10
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- The archives of bone and joint surgery
- Publication Type :
- Academic Journal
- Accession number :
- 35755787
- Full Text :
- https://doi.org/10.22038/ABJS.2021.58351.2889