Back to Search
Start Over
Monopolar, bipolar, tripolar, and tetrapolar configurations in robot assisted electrical impedance sensing.
- Source :
-
Biomedical physics & engineering express [Biomed Phys Eng Express] 2022 Jul 19; Vol. 8 (5). Date of Electronic Publication: 2022 Jul 19. - Publication Year :
- 2022
-
Abstract
- Objective. Tissue recognition is a critical process during a Robot-assisted minimally invasive surgery (RMIS) and it relies on the involvement of advanced sensing technology. Approach. In this paper, the concept of Robot Assisted Electrical Impedance Sensing (RAEIS) is utilized and further developed aiming to sense the electrical bioimpedance of target tissue directly based on the existing robotic instruments and control strategy. Specifically, we present a new sensing configuration called pseudo-tetrapolar method. With the help of robotic control, we can achieve a similar configuration as traditional tetrapolar, and with better accuracy. Main results. Five configurations including monopolar, bipolar, tripolar, tetrapolar and pseudo-tetrapolar are analyzed and compared through simulation experiments. Advantages and disadvantages of each configuration are thus discussed. Significance. This study investigates the measurement of tissue electrical property directly based on the existing robotic surgical instruments. Specifically, different sensing configurations can be realized through different connection and control strategies, making them suitable for different application scenarios.<br /> (© 2022 IOP Publishing Ltd.)
- Subjects :
- Electric Impedance
Robotic Surgical Procedures
Robotics methods
Subjects
Details
- Language :
- English
- ISSN :
- 2057-1976
- Volume :
- 8
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Biomedical physics & engineering express
- Publication Type :
- Academic Journal
- Accession number :
- 35728560
- Full Text :
- https://doi.org/10.1088/2057-1976/ac7adb