Back to Search Start Over

Hierarchical Multi-Resolution Graph-Cuts for Water-Fat-Silicone Separation in Breast MRI.

Authors :
Stelter JK
Boehm C
Ruschke S
Weiss K
Diefenbach MN
Wu M
Borde T
Schmidt GP
Makowski MR
Fallenberg EM
Karampinos DC
Source :
IEEE transactions on medical imaging [IEEE Trans Med Imaging] 2022 Nov; Vol. 41 (11), pp. 3253-3265. Date of Electronic Publication: 2022 Oct 27.
Publication Year :
2022

Abstract

Water-fat separation is a non-linear non-convex parameter estimation problem in magnetic resonance imaging typically solved using spatial constraints. However, there is still limited knowledge on how to separate in vivo three chemical species in the presence of magnetic field inhomogeneities. The proposed method uses multiple graph-cuts in a hierarchical multi-resolution framework to perform robust chemical species separation in the breast for subjects with and without silicone implants. Experimental results show that the proposed method can decrease the computational time for water-fat separation and perform accurate water-fat-silicone separation with only a limited number of acquired echo images at 3 T. The silicone-separated images have an improved spatial resolution and image contrast compared to conventional scans used for regular monitoring of the silicone implant's integrity.

Details

Language :
English
ISSN :
1558-254X
Volume :
41
Issue :
11
Database :
MEDLINE
Journal :
IEEE transactions on medical imaging
Publication Type :
Academic Journal
Accession number :
35657831
Full Text :
https://doi.org/10.1109/TMI.2022.3180302