Easy and computer-time-saving implementation of the van der Pauw method including anisotropy and probe positioning correction factors using approximate closed-form analytical functions.

Authors :
Hurand S
Chommaux T
Renault PO
Girardeau T
Paumier F
Source :
The Review of scientific instruments [Rev Sci Instrum] 2022 May 01; Vol. 93 (5), pp. 053907.
Publication Year :
2022

Abstract

Measurements in a van der Pauw configuration often require the use of complicated mathematical expressions or implicit relations, which may be computer-time-consuming or tedious to implement in a given software. Thus, a closed-form expression is often desirable. We propose to approximate these relations by closed-form analytical functions using only basic operators that can be easily implemented. We present explicitly the functions and the numerical values of their parameters for the cases of standard van der Pauw measurement, anisotropic resistivities, and the probe position correction factor. The discrepancy with the exact values is smaller than 10 <superscript>-5</superscript> , which is sufficient for practical purposes.

Details

Language :
English
ISSN :
1089-7623
Volume :
93
Issue :
5
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
35649772
Full Text :
https://doi.org/10.1063/5.0068682