Back to Search
Start Over
A Simple Technique for Reliable Donor Organ Temperature Management With the SherpaPak System.
- Source :
-
ASAIO journal (American Society for Artificial Internal Organs : 1992) [ASAIO J] 2022 Aug 01; Vol. 68 (8), pp. e134-e135. Date of Electronic Publication: 2022 Jan 19. - Publication Year :
- 2022
-
Abstract
- Here, we present our technique for thermoregulation of cardiac preservation solution with the SherpaPak device before heart transplantation. If it is above the ideal range (5-10°C) to avoid organ injury and ensure proper function, the cardiac preservation solution bag should be placed in ice water to bring down temperature until the ideal range is achieved. SherpaPak has limitation in immediately correcting initial temperatures outside of the normal range, so we believe this simple step should be included into standard SherpaPak procedure to ensure its effectiveness and to preserve the best organ function.<br />Competing Interests: Disclosure: The authors have no funding and conflicts of interest to report.<br /> (Copyright © ASAIO 2022.)
Details
- Language :
- English
- ISSN :
- 1538-943X
- Volume :
- 68
- Issue :
- 8
- Database :
- MEDLINE
- Journal :
- ASAIO journal (American Society for Artificial Internal Organs : 1992)
- Publication Type :
- Academic Journal
- Accession number :
- 35605255
- Full Text :
- https://doi.org/10.1097/MAT.0000000000001657