Back to Search Start Over

Conversion of imager-specific response to tissue phantom fluorescence into system of units-traceable units.

Authors :
Litorja M
Source :
Journal of biomedical optics [J Biomed Opt] 2022 May; Vol. 27 (7).
Publication Year :
2022

Abstract

Significance: The fluorescence-guided imaging for surgical intervention community recognizes the need for performance standards for these imaging devices. Tissue phantoms are used to track an imager's performance as a fluorescence detector, but imager-specific units are of limited utility.<br />Aim: Tissue phantoms can be calibrated to be traceable to the international system of units (SI) and in turn be used to calibrate imagers such that fluorescence measurements can be reported in universally accepted units.<br />Approach: The radiometry to convert imager-specific arbitrary digital counts to SI-traceable unit of watts is described in this paper.<br />Results: An example of an imager calibration is included.<br />Conclusions: Calibrated tissue phantoms become a tool for metrological traceability.

Details

Language :
English
ISSN :
1560-2281
Volume :
27
Issue :
7
Database :
MEDLINE
Journal :
Journal of biomedical optics
Publication Type :
Academic Journal
Accession number :
35552461
Full Text :
https://doi.org/10.1117/1.JBO.27.7.074715