Back to Search Start Over

Increasing the transmission efficiency of transcranial ultrasound using a dual-mode conversion technique based on Lamb waves.

Authors :
Kang KC
Kim YH
Kim JN
Kabir M
Zhang Y
Ghanouni P
Park KK
Firouzi K
Khuri-Yakub BT
Source :
The Journal of the Acoustical Society of America [J Acoust Soc Am] 2022 Mar; Vol. 151 (3), pp. 2159.
Publication Year :
2022

Abstract

Transcranial focused ultrasound (FUS) is a noninvasive treatment for brain tumors and neuromodulation. Based on normal incidence, conventional FUS techniques use a focused or an array of ultrasonic transducers to overcome the attenuation and absorption of ultrasound in the skull; however, this remains the main limitation of using FUS. A dual-mode conversion technique based on Lamb waves is proposed to achieve high transmission efficiency. This concept was validated using the finite element analysis (FEA) and experiments based on changes in the incident angle. Aluminum, plexiglass, and a human skull were used as materials with different attenuations. The transmission loss was calculated for each material, and the results were compared with the reflectance function of the Lamb waves. Oblique incidence based on dual-mode conversion exhibited a better transmission efficiency than that of a normal incidence for all of the specimens. The total transmission losses for the materials were 13.7, 15.46, and 3.91 dB less than those associated with the normal incidence. A wedge transducer was designed and fabricated to implement the proposed method. The results demonstrated the potential applicability of the dual-mode conversion technique for the human skull.

Details

Language :
English
ISSN :
1520-8524
Volume :
151
Issue :
3
Database :
MEDLINE
Journal :
The Journal of the Acoustical Society of America
Publication Type :
Academic Journal
Accession number :
35364946
Full Text :
https://doi.org/10.1121/10.0009849