Development of a one-step analysis method for several amino acids using a microfluidic paper-based analytical device.
- Source :
-
Scientific reports [Sci Rep] 2022 Mar 02; Vol. 12 (1), pp. 3427. Date of Electronic Publication: 2022 Mar 02. - Publication Year :
- 2022
-
Abstract
- A one-step analysis method was developed for four types of amino acids using a microfluidic paper-based analytical device fabricated from chromatography filtration paper and laminate films. Aminoacyl-tRNA synthetase was used to detect each amino acid. The obtained laminated paper-based analytical device (LPAD) contained four enzymatic reaction areas. Colorimetric detection was performed based on the molybdenum blue reaction. A model method for the simple, easy, and simultaneous detection of several amino acid concentrations was suggested, in contrast to the conventional methods such as HPLC or LC-MS. The method provided a selective quantification at the ranges of 3.6-100 μM for tryptophan, 10.1-100 μM for glycine, 5.9-100 μM for histidine and 5.6-100 μM for lysine with a detection limit of 1.1 μM, 3.3 μM, 1.9 μM and 1.8 μM, respectively. LPAD fabrication was considerably simple, and the subsequent detection process was easy and required a short period of time (within 15 min).<br /> (© 2022. The Author(s).)