Automated Planning for Prostate Stereotactic Body Radiation Therapy on the 1.5 T MR-Linac.
- Source :
-
Advances in radiation oncology [Adv Radiat Oncol] 2022 Feb 12; Vol. 7 (3), pp. 100865. Date of Electronic Publication: 2022 Feb 12 (Print Publication: 2022). - Publication Year :
- 2022
-
Abstract
- Purpose: Adaptive stereotactic body radiation therapy (SBRT) for prostate cancer (PC) by the 1.5 T MR-linac currently requires online planning by an expert user. A fully automated and user-independent solution to adaptive planning (mCycle) of PC-SBRT was compared with user's plans for the 1.5 T MR-linac.<br />Methods and Materials: Fifty adapted plans on daily magnetic resonance imaging scans for 10 patients with PC treated by 35 Gy (prescription dose [D <subscript>p</subscript> ]) in 5 fractions were reoptimized offline from scratch, both by an expert planner (manual) and by mCycle. Manual plans consisted of multicriterial optimization (MCO) of the fluence map plus manual tweaking in segmentation, whereas in mCycle plans, the objectives were sequentially optimized by MCO according to an a-priori assigned priority list. The main criteria for planning approval were a dose ≥95% of the D <subscript>p</subscript> to at least 95% of the planning target volume (PTV), V <subscript>33.2</subscript> (PTV) ≥ 95%, a dose less than the D <subscript>p</subscript> to the hottest cubic centimeter (V <subscript>35</subscript> ≤ 1 cm <superscript>3</superscript> ) of rectum, bladder, penile bulb, and urethral planning risk volume (ie, urethra plus 3 mm isotropically), and V <subscript>32</subscript> ≤ 5%, V <subscript>28</subscript> ≤ 10%, and V <subscript>18</subscript> ≤ 35% to the rectum. Such dose-volume metrics, plus some efficiency and deliverability metrics, were used for the comparison of mCycle versus manual plans.<br />Results: mCycle plans improved target dose coverage, with V <subscript>33.2</subscript> (PTV) passing on average (±1 SD) from 95.7% (±1.0%) for manual plans to 97.5% (±1.3%) for mCycle plans ( P < .001), and rectal dose sparing, with significantly reduced V <subscript>32</subscript> , V <subscript>28</subscript> , and V <subscript>18</subscript> ( P ≤ .004). Although at an equivalent number of segments, mCycle plans consumed moderately more monitor units (+17%) and delivery time (+9%) ( P < .001), whereas they were generally faster (-19%) in terms of optimization times ( P < .019). No significant differences were found for the passing rates of locally normalized γ (3 mm, 3%) ( P  = .059) and γ (2 mm, 2%) ( P  = .432) deliverability metrics.<br />Conclusions: In the offline setting, mCycle proved to be a trustable solution for automated planning of PC-SBRT on the 1.5 T MR-linac. mCycle integration in the online workflow will free the user from the challenging online-optimization task.<br /> (© 2021 The Author(s).)
Details
- Language :
- English
- ISSN :
- 2452-1094
- Volume :
- 7
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Advances in radiation oncology
- Publication Type :
- Academic Journal
- Accession number :
- 35198836
- Full Text :
- https://doi.org/10.1016/j.adro.2021.100865