Tests of the accuracy of a data reduction method for determination of acoustic backscatter coefficients.

Authors :
Insana MF
Madsen EL
Hall TJ
Zagzebski JA
Source :
The Journal of the Acoustical Society of America [J Acoust Soc Am] 1986 May; Vol. 79 (5), pp. 1230-6.
Publication Year :
1986

Abstract

The accuracy of a new method for measuring ultrasonic backscatter coefficients was tested, using narrow-band pulses and well-defined media having scatterers randomly distributed in space. Experimentally determined values agree very well with theoretical values for wide ranges of experimental parameters, these ranges being applicable in measurements made on human soft tissues. An important outcome is that the method yields accurate results for scattering media positioned anywhere from the nearfield through the farfield of the nonfocused transducers employed. In addition, backscatter coefficients can be determined for a broad range of gate durations.

Details

Language :
English
ISSN :
0001-4966
Volume :
79
Issue :
5
Database :
MEDLINE
Journal :
The Journal of the Acoustical Society of America
Publication Type :
Academic Journal
Accession number :
3519721
Full Text :
https://doi.org/10.1121/1.393702