Comparison of Phase Estimation Methods for Quantitative Susceptibility Mapping Using a Rotating-Tube Phantom.
- Source :
-
Radiology research and practice [Radiol Res Pract] 2021 Nov 24; Vol. 2021, pp. 1898461. Date of Electronic Publication: 2021 Nov 24 (Print Publication: 2021). - Publication Year :
- 2021
-
Abstract
- Quantitative Susceptibility Mapping (QSM) is an MRI tool with the potential to reveal pathological changes from magnetic susceptibility measurements. Before phase data can be used to recover susceptibility (Δ χ ), the QSM process begins with two steps: data acquisition and phase estimation. We assess the performance of these steps, when applied without user intervention, on several variations of a phantom imaging task. We used a rotating-tube phantom with five tubes ranging from Δ χ =0.05 ppm to Δ χ =0.336 ppm. MRI data was acquired at nine angles of rotation for four different pulse sequences. The images were processed by 10 phase estimation algorithms including Laplacian, region-growing, branch-cut, temporal unwrapping, and maximum-likelihood methods, resulting in approximately 90 different combinations of data acquisition and phase estimation methods. We analyzed errors between measured and expected phases using the probability mass function and Cumulative Distribution Function. Repeatable acquisition and estimation methods were identified based on the probability of relative phase errors. For single-echo GRE and segmented EPI sequences, a region-growing method was most reliable with Pr (relative error <0.1) = 0.95 and 0.90, respectively. For multiecho sequences, a maximum-likelihood method was most reliable with Pr (relative error <0.1) = 0.97. The most repeatable multiecho methods outperformed the most repeatable single-echo methods. We found a wide range of repeatability and reproducibility for off-the-shelf MRI acquisition and phase estimation approaches, and this variability may prevent the techniques from being widely integrated in clinical workflows. The error was dominated in many cases by spatially discontinuous phase unwrapping errors. Any postprocessing applied on erroneous phase estimates, such as QSM's background field removal and dipole inversion, would suffer from error propagation. Our paradigm identifies methods that yield consistent and accurate phase estimates that would ultimately yield consistent and accurate Δ χ estimates.<br />Competing Interests: The authors K. E.K, B. P.B, S. C., S. E.R., W.-T.W., J. A.B, and D. L.P. have no conflicts of interest. J. D. is the inventor of the MAGPI method and is the founder of MAGPI LLC.<br /> (Copyright © 2021 Kathryn E. Keenan et al.)
Details
- Language :
- English
- ISSN :
- 2090-1941
- Volume :
- 2021
- Database :
- MEDLINE
- Journal :
- Radiology research and practice
- Publication Type :
- Academic Journal
- Accession number :
- 34868681
- Full Text :
- https://doi.org/10.1155/2021/1898461