Back to Search Start Over

[Research on Reliability Index and Realization of Magnetic Resonance (MR) Based on Clinical Use Condition].

Authors :
Jia D
He F
Jin W
Zhang W
Liu Y
Zou J
Source :
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation [Zhongguo Yi Liao Qi Xie Za Zhi] 2021 Nov 30; Vol. 45 (6), pp. 628-635.
Publication Year :
2021

Abstract

Combined with the clinical use condition of MR in use in Shanghai Sixth People's Hospital, MR components are divided into scanning type I and scanning type II. At the same time, combined with the main loss force of MR components, the research divides MR components into dynamic components and electric thermal components. In this study, a complete set of MR system reliability indexes and implementation methods are given, including system reliability index determination, system reliability allocation, component reliability index realization, system reliability prediction and system reliability verification. At the same time, this study also gives the methods of reliability prediction and reliability verification, and gives the MTBF calculation method of MR system based on clinical use data statistics.

Details

Language :
Chinese
ISSN :
1671-7104
Volume :
45
Issue :
6
Database :
MEDLINE
Journal :
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Publication Type :
Academic Journal
Accession number :
34862775
Full Text :
https://doi.org/10.3969/j.issn.1671-7104.2021.06.009