The use of BCYE medium for the detection of Legionella in environmental water samples: an appropriate update to ISO 11731:2017 standard?
- Source :
-
Diagnostic microbiology and infectious disease [Diagn Microbiol Infect Dis] 2022 Feb; Vol. 102 (2), pp. 115593. Date of Electronic Publication: 2021 Nov 07. - Publication Year :
- 2022
-
Abstract
- We evaluated the diagnostic performances of 2 media (BCYE, MWY) on 951 Legionella-positive hospital water samples. MWY allowed detecting Legionella in 89.2% of samples, but in 10.8% (103/951) Legionella was found on BCYE plates only. In samples where Legionella was isolated with other microorganisms (663/951), MWY was essential to detect the majority of positive samples (349/663, 52.6%), as fewer plates resulted unreadable; however, in those containing Legionella only, a higher frequency of positive samples was recorded with BCYE (94.8%, 273/288) compared to MWY (85.1%, 245/288). Considering the 484 concordant positive samples, overall Legionella counts were significantly higher on BCYE (P = 0.0029), with 47% of samples showing higher counts on BCYE compared to MWY plates. Furthermore, discordant samples (positive on only one medium) showed different relative proportions between Legionella pneumophila and non-pneumophila, the latter being found more frequently on BCYE only (P = 0.0296).Our findings confirm the appropriateness of the ISO 11731:2017 update.<br />Competing Interests: Declaration of competing interests The authors report no conflicts of interest relevant to this article.<br /> (Copyright © 2021 Elsevier Inc. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1879-0070
- Volume :
- 102
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Diagnostic microbiology and infectious disease
- Publication Type :
- Academic Journal
- Accession number :
- 34852987
- Full Text :
- https://doi.org/10.1016/j.diagmicrobio.2021.115593