Back to Search
Start Over
Diagnostic Accuracy of Point-of-Care Tests Measuring Glycosylated Haemoglobin (HbA1c) for Glycemic Control: A Field Study in India.
- Source :
-
Cureus [Cureus] 2021 Sep 13; Vol. 13 (9), pp. e17920. Date of Electronic Publication: 2021 Sep 13 (Print Publication: 2021). - Publication Year :
- 2021
-
Abstract
- Objectives This study was performed to estimate diagnostic accuracy of the two commercially available point-of-care tests to identify poor glycemic control defined by HbA1c levels, with high-performance liquid chromatography (HPLC) as a reference. Settings The study was carried at two locations, general medical outpatient department of a teaching medical college in Bhopal (urban), and a primary health care centre in a rural area in the state of Madhya Pradesh, India. Participants All individuals with diabetes mellitus who presented to the health care facility for assessment of glycemic control. We compared HbA1c estimated from two index tests (Hemocue Hb 501, Sweden; SD Biosensor, South Korea) from capillary blood samples with HPLC performed from venous blood, as a reference standard. Primary and secondary outcome measures Diagnostic properties of index tests such as sensitivity, specificity, positive and negative predictive value and diagnostic accuracy for identifying poor glycemic control were primary outcome measures. Lin's concordance correlation coefficient (CCC) was secondary outcome measure. Results Out of 114 patients, all received reference standard - 103 patients received Hemocue A1C test and 110 patients received SD Biosensor test. Overall both the index tests had similar diagnostic accuracy estimates. The area under the Receiver Operating Curve for SDA1c device was 0.935 (95% CI 0.886-0.983), and for Hemocue device was 0.938 (95% CI 0.893-0.984). The Hemocue device HbA1c value of above 7.0 (positive) correctly predicted poor glycemic control 92% times (81.58% for SD device). There were 4 vs. 11 device failures and 14 vs. 12 failures with SD and Hemocue, respectively. Ambient air temperatures were no different for the device test failures. Conclusions Commercially available point-of-care tests evaluated in this study are comparable and an acceptable alternative to HPLC-based measurements for the assessment of glycemic control. Tests and device failure rates of both the index tests are similar.<br />Competing Interests: The authors have declared that no competing interests exist.<br /> (Copyright © 2021, Khadanga et al.)
Details
- Language :
- English
- ISSN :
- 2168-8184
- Volume :
- 13
- Issue :
- 9
- Database :
- MEDLINE
- Journal :
- Cureus
- Publication Type :
- Academic Journal
- Accession number :
- 34660113
- Full Text :
- https://doi.org/10.7759/cureus.17920