Back to Search Start Over

IMAGE QUALITY STUDY OF RADIATION-REDUCED COMBINED CHEST AND ABDOMEN/PELVIS CT COMPARED WITH A STANDARD PROTOCOL.

Authors :
Bos D
Zensen S
Opitz M
Haubold J
Forsting M
Nassenstein K
Guberina N
Wetter A
Source :
Radiation protection dosimetry [Radiat Prot Dosimetry] 2021 Nov 12; Vol. 196 (3-4), pp. 190-198.
Publication Year :
2021

Abstract

The purpose of this study is to examine radiation doses and image quality of a low-dose (LD) protocol for chest and abdomen/pelvis (CAP) CT compared with a standard (STD) protocol. A total of 361 patients were included between October 2019 and April 2020; 104 patients with LD-protocol (100 kV, ref mAs 80 (chest)/145 (abdomen/pelvis)) and 257 patients with STD-protocol (100 kV, ref mAs 100 (chest)/180 (abdomen/pelvis)) at second-generation dual-source CT. Radiation doses for CTDIvol and DLP, and objective and subjective image qualities of 50 examinations from each group were evaluated. The LD-protocol applied significantly lower radiation doses compared with the STD-protocol (p < 0.001), achieving a dose reduction by 37% for the median DLP in chest, 19% in abdomen/pelvis and 22% in total. Median total DLP was 342 mGy·cm (LD) vs. 436 mGy·cm (STD). The LD-CAP CT protocol achieved a significant dose reduction far below national diagnostic reference levels, ensuring acceptable and good image quality.<br /> (© The Author(s) 2021. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1742-3406
Volume :
196
Issue :
3-4
Database :
MEDLINE
Journal :
Radiation protection dosimetry
Publication Type :
Academic Journal
Accession number :
34635920
Full Text :
https://doi.org/10.1093/rpd/ncab147