Back to Search Start Over

Computer-assisted photoreceptor assessment on Heidelberg Engineering Spectralis™ High Magnification Module™ images.

Authors :
Mulders TWF
Klevering BJ
Hoyng CB
Theelen T
Source :
Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie [Graefes Arch Clin Exp Ophthalmol] 2021 Nov; Vol. 259 (11), pp. 3311-3320. Date of Electronic Publication: 2021 Aug 06.
Publication Year :
2021

Abstract

Purpose: To evaluate reliability and repeatability of computer-assisted measurements of cone photoreceptor metrics on Heidelberg Engineering Spectralis™ High Magnification Module (HMM™) Automatic Real-time Tracking (ART™) images.<br />Methods: We analyzed HMM™ images in three separate study arms. Computer-assisted cone identification software was validated using an open-access adaptive optics (AO) dataset. We compared results of the first arm to data from AO and histology. We evaluated intersession repeatability of our computer-assisted cone analysis in the second arm. We assessed the capability of HMM™ to visualize cones in the presence of pathology in the third arm.<br />Results: We included 10 healthy subjects in the first arm of our study, 5 additional healthy participants in the second arm and 5 patients in the third arm. In total, we analyzed 225 regions of interest on HMM™ images. We were able to automatically identify cone photoreceptors and assess corresponding metrics at all eccentricities between 2 and 9° from the fovea. Cone density significantly declined with increasing eccentricity (p = 4.890E-26, Friedman test). With increasing eccentricity, we found a significant increase in intercell distance (p = 2.196E-25, Friedman test) and nearest neighbor distance (p = 1.997E-25, Friedman test). Cone hexagonality ranged between 71 and 85%. We found excellent automated intersession repeatability of cone density counts and spacing measurements. In pathology, we were also able to repeatedly visualize photoreceptors.<br />Conclusion: Computer-assisted cone photoreceptor analysis on Spectralis™ HMM™ images is feasible, and most cone metrics show excellent repeatability. HMM™ imaging may be useful for photoreceptor analysis as progression marker in outer retinal disease.<br /> (© 2021. The Author(s).)

Details

Language :
English
ISSN :
1435-702X
Volume :
259
Issue :
11
Database :
MEDLINE
Journal :
Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie
Publication Type :
Academic Journal
Accession number :
34357417
Full Text :
https://doi.org/10.1007/s00417-021-05326-6