Intrafraction stability using full head mask for brain stereotactic radiotherapy.

Authors :
Tomihara J
Takatsu J
Sugimoto S
Shikama N
Sasai K
Source :
Journal of applied clinical medical physics [J Appl Clin Med Phys] 2021 Sep; Vol. 22 (9), pp. 360-370. Date of Electronic Publication: 2021 Aug 04.
Publication Year :
2021

Abstract

Purpose: We investigated the immobilization accuracy of a new type of thermoplastic mask-the Double Shell Positioning System (DSPS)-in terms of geometry and dose delivery.<br />Methods: Thirty-one consecutive patients with 1-5 brain metastases treated with stereotactic radiotherapy (SRT) were selected and divided into two groups. Patients were divided into two groups. One group of patients was immobilized by the DSPS (n = 9). Another group of patients was immobilized by a combination of the DSPS and a mouthpiece (n = 22). Patient repositioning was performed with cone beam computed tomography (CBCT) and six-degree of freedom couch. Additionally, CBCT images were acquired before and after treatment. Registration errors were analyzed with off-line review. The inter- and intrafractional setup errors, and planning target volume (PTV) margin were also calculated. Delivered doses were calculated by shifting the isocenter according to inter- and intrafractional setup errors. Dose differences of GTV D <subscript>99%</subscript> were compared between planned and delivered doses against the modified PTV margin of 1 mm.<br />Results: Interfractional setup errors associated with the mouthpiece group were significantly smaller than the translation errors in another group (p = 0.03). Intrafractional setup errors for the two groups were almost the same in all directions. PTV margins were 0.89 mm, 0.75 mm, and 0.90 mm for the DSPS combined with the mouthpiece in lateral, vertical, and longitudinal directions, respectively. Similarly, PTV margins were 1.20 mm, 0.72 mm, and 1.37 mm for the DSPS in the lateral, vertical, and longitudinal directions, respectively. Dose differences between planned and delivered doses were small enough to be within 1% for both groups.<br />Conclusions: The geometric and dosimetric assessments revealed that the DSPS provides sufficient immobilization accuracy. Higher accuracy can be expected when the immobilization is combined with the use of a mouthpiece.<br /> (© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.)

Details

Language :
English
ISSN :
1526-9914
Volume :
22
Issue :
9
Database :
MEDLINE
Journal :
Journal of applied clinical medical physics
Publication Type :
Academic Journal
Accession number :
34347933
Full Text :
https://doi.org/10.1002/acm2.13382