Back to Search Start Over

Efficacy and safety of the GOLD FORCE multicentre randomized clinical trial: multielectrode phased radiofrequency vs. irrigated radiofrequency single-tip catheter with contact force ablation for treatment of symptomatic paroxysmal atrial fibrillation.

Authors :
Wintgens LIS
Klaver MN
Maarse M
Spitzer SG
Langbein A
Swaans MJ
Van Dijk VF
Balt JC
Wijffels MCEF
Tijssen JGP
Elvan A
Boersma LVA
Source :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2021 Dec 07; Vol. 23 (12), pp. 1931-1938.
Publication Year :
2021

Abstract

Aims: Pulmonary vein isolation (PVI) for atrial fibrillation (AF) has become increasingly safe and effective with the evolution of single-tip ablation catheters aided by contact force sensing (ST-CF) and single-shot devices such as the second-generation pulmonary vein ablation catheter (PVAC) Gold multi-electrode array. The multicentre randomized GOLD FORCE trial was conducted to evaluate non-inferiority of safety and efficacy of PVAC Gold PVI compared to ST-CF ablation for paroxysmal AF.<br />Methods and Results: The primary efficacy endpoint documented AF recurrence ≥30 s was assessed by time-to-first-event analysis after a 90-day blanking period using repeated 7-day Holters. Secondary endpoints include acute success and procedural characteristics. Safety endpoints included procedural complications, stroke/transient ischaemic attack (TIA), tamponade, bleeding, and access site complications. Two hundred and eight patients underwent randomization and PVI (103 assigned to PVAC Gold, 105 to ST-CF). Acute success rates were 95% and 97% for PVAC Gold and ST-CF, respectively. At 12 months, AF recurrence was observed in 46.6% of the PVAC Gold group and in 26.2% of the ST-CF group [absolute efficacy difference 20.4% (95% confidence interval, CI 7.5-33.2%), hazard ratio 2.05 (95% CI 1.28-3.29), P = 0.003]. PVAC Gold had significantly shorter procedure and ablation times. Complication rates were 5.7% and 4.9% for PVAC Gold and ST-CF, respectively (P = 0.782).<br />Conclusion: In this multicentre randomized clinical trial, ablation with ST-CF and PVAC Gold ablation catheters non-inferiority for efficacy was not met. AF recurrence was significantly more frequent in the PVAC Gold group compared to single-tip contact force group. Both groups had similarly low rates of adverse events. PVAC Gold ablation had significantly shorter procedure and ablation times.<br /> (Published on behalf of the European Society of Cardiology. All rights reserved. © The Author(s) 2021. For permissions, please email: journals.permissions@oup.com.)

Details

Language :
English
ISSN :
1532-2092
Volume :
23
Issue :
12
Database :
MEDLINE
Journal :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
34279627
Full Text :
https://doi.org/10.1093/europace/euab168