A Comparative Study of Multidetector Computed Tomography, Cone Beam Computed Tomography, and Computed Microtomography on Trabecular Bone Structures in the Human Mandible: An Ex Vivo Study.

Authors :
Sarmento VA
de Oliveira Gonzalez TFL
Lopes RT
Novaes M
Borges SM
Rubira-Bullen IRF
Source :
Journal of computer assisted tomography [J Comput Assist Tomogr] 2021 Jul-Aug 01; Vol. 45 (4), pp. 552-556.
Publication Year :
2021

Abstract

Objective: The aim of this study was to evaluate the accuracy of multislice computed tomography (MSCT) and cone beam computed tomography (CBCT) scanners in the assessment of trabecular bone microarchitecture by radiomorphometric analysis.<br />Methods: Eight dry human hemimandibles were selected and scanned by 4- and 128-MSCT, CBCT with different voxel sizes, and a computed microtomography (micro-CT). The images were spatially aligned, the axial slices were segmented using CTAn software (Skyscan, Konitch, Belgium), and 10 bone morphometric parameters were quantified. The micro-CT images were the criterion standard.<br />Results: The results show the superiority of CBCT images with 0.2-mm voxel size and 128-MSCT.<br />Conclusions: We can conclude that the CBCT scanner with 0.2-mm voxel size and 128-MSCT had better performance on reproducibility of the distribution and organization of trabecular bone and marrow spaces evaluated in this study.<br />Competing Interests: The authors declare no conflict of interest.<br /> (Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved.)

Details

Language :
English
ISSN :
1532-3145
Volume :
45
Issue :
4
Database :
MEDLINE
Journal :
Journal of computer assisted tomography
Publication Type :
Academic Journal
Accession number :
34270480
Full Text :
https://doi.org/10.1097/RCT.0000000000001191