- The goal of developing a PET dedicated to the brain (CareMiBrain) has evolved from its initial approach to diagnosis and monitoring of dementias, to the more ambitious of creating a revolutionary clinical pathway for the knowledge and personalized treatment of multiple neurological diseases. The main innovative feature of CareMiBrain is the use of detectors with continuous crystals, which allow a high resolution determination of the depth of annihilation photons interaction within the thickness of the scintillation crystal. The technical validation phase of the equipment consisted of a pilot, prospective and observational study whose objective was to obtain the first images (40 patients), analyze them and make adjustments in the acquisition, reconstruction and correction parameters, comparing the image quality of the CareMiBrain equipment with that of the whole-body PET/CT. Thanks to the team meetings and the joint analysis of the images, it was possible to detect its weak points and some of its causes. The calibration, acquisition and processing processes, as well as the reconstruction, were optimized, the number of iterations was set to achieve the best signal-to-noise ratio, the random correction was optimized and a post-processing algorithm was included in the reconstruction algorithm. The main technical improvements implemented in this phase of technical validation carried out through collaboration of the Services of Nuclear Medicine and Neurology of the Hospital Clínico San Carlos with the Spanish company Oncovision will be exposed in a project financed with funds from the European Union (Horizon 2020 innovation program, 713323).<br /> (Copyright © 2021 Sociedad Española de Medicina Nuclear e Imagen Molecular. Published by Elsevier España, S.L.U. All rights reserved.)
- Subjects :
- Adult
Calibration
Dementia diagnostic imaging
Equipment Design
Humans
Pilot Projects
Positron Emission Tomography Computed Tomography
Positron-Emission Tomography instrumentation
Positron-Emission Tomography standards
Prospective Studies
Signal-To-Noise Ratio
Whole Body Imaging methods
Whole Body Imaging standards
Brain diagnostic imaging
Neurodegenerative Diseases diagnostic imaging
Positron-Emission Tomography methods
Quality Improvement
Subjects
Details
- Language :
- English
- ISSN :
- 2253-8089
- Volume :
- 40
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Revista espanola de medicina nuclear e imagen molecular
- Publication Type :
- Academic Journal
- Accession number :
- 34218886
- Full Text :
- https://doi.org/10.1016/j.remnie.2021.05.001