Back to Search Start Over

Experimental determination of cascade parameters of a hearing-aid microphone via the two-load method.

Authors :
Egolf DP
Haley BT
Bauer KM
Howell HC
Larson VD
Source :
The Journal of the Acoustical Society of America [J Acoust Soc Am] 1988 Jun; Vol. 83 (6), pp. 2439-46.
Publication Year :
1988

Abstract

Presented in this article is a computer-aided experimental method for obtaining the cascade parameters of the two-port model of a miniature hearing-aid microphone. The method is an adaptation of the "two-load" method [D.P. Egolf and R.G. Leonard, J. Acoust. Soc. Am. 62, 1013-1023 (1977)] to acoustoelectric, rather than electroacoustic, transducers. The cascade parameters of a particular microphone, determined by this method, were within 2.5 dB of the manufacturer's published open-circuit sensitivity data. In an attempt to further verify the numerical cascade-parameter data, a two-port model of the microphone was used to simulate experimental voltages developed across two different complex electrical load impedances attached to the microphone. The results showed experimental/simulation differences of no greater than 3.0 dB at any frequency. The two-port microphone model and associated cascade parameters are currently being incorporated into a computer-based plan for mathematical simulation of an entire in situ hearing aid.

Details

Language :
English
ISSN :
0001-4966
Volume :
83
Issue :
6
Database :
MEDLINE
Journal :
The Journal of the Acoustical Society of America
Publication Type :
Academic Journal
Accession number :
3411033
Full Text :
https://doi.org/10.1121/1.396323