Evaluation of the Fully Automated Chemiluminescence Analyzer Liaison XL for the Performance of the QuantiFERON-TB Gold Plus Assay in an Area with a Low Incidence of Tuberculosis.

Authors :
Fernández-Huerta M
Moreto C
Vila-Olmo N
García de Cara EI
Basaez MC
Santín M
Alcaide F
Source :
Journal of clinical microbiology [J Clin Microbiol] 2021 Jul 19; Vol. 59 (8), pp. e0060321. Date of Electronic Publication: 2021 Jul 19.
Publication Year :
2021

Abstract

Diagnosis of latent tuberculosis infection (LTBI) is considered key in the control of tuberculosis. Interferon gamma (IFN-γ) release assays, such as the QuantiFERON-TB Gold Plus test (QFT-Plus), are now widely implemented for the in vitro diagnosis of LTBI. To date, the detection and quantification of IFN-γ has been mostly performed with semiautomated enzyme-linked immunosorbent assays (ELISAs), but several limitations currently exist. The study aims to evaluate the chemiluminescence immunoassay (CLIA) analyzer Liaison XL compared to ELISA for the performance of the QFT-Plus test. Between February and April 2020, 333 heparin blood samples from 323 adult patients were collected at a tertiary teaching hospital in Barcelona, Spain. Overall, the CLIA analyzer Liaison XL performed well for the detection of IFN-γ compared to the ELISA method, demonstrating substantial agreement (κ, 0.872) and great correlation between assays (r, >0.950). CLIA produced significantly higher values of IFN-γ IU per milliliter than the ELISA ( P  = 0.004 for the TB1 tube and P  = 0.010 for the TB2 tube). Many discrepant cases (8/15, 53.3%) corresponded to indeterminate results with ELISA (NIL-corrected mitogen value of <0.5 IU/ml), which, when analyzed with the CLIA analyzer Liaison XL, reverted to interpretable results. In conclusion, this analysis suggests that CLIA presents a greater sensitivity for the identification of LTBI, especially among immunocompromised patients. Furthermore, the analytical variability reported between both ELISA and CLIA methods, especially around the standardized 0.35-IU/ml positivity threshold, suggests the need to refine the interpretative algorithm.

Details

Language :
English
ISSN :
1098-660X
Volume :
59
Issue :
8
Database :
MEDLINE
Journal :
Journal of clinical microbiology
Publication Type :
Academic Journal
Accession number :
34076474
Full Text :
https://doi.org/10.1128/JCM.00603-21